Digital Breast Tomosynthesis QC Phantom

Digital Breast Tomosynthesis QC Phantom

The Digital Breast Tomosynthesis QC Phantom supports comprehensive testing of digital breast tomosynthesis systems.
WORKFLOW STEP
TASK GROUP

Overview

The Modular DBT Phantom is no longer being manufactured or sold by Sun Nuclear. We are pleased to offer the following suitable alternative from CIRS.

Assess Breast Tomosynthesis Image Quality
The CIRS Digital Breast Tomosynthesis (DBT) QC Phantom is designed to address quality control for DBT systems. The phantom can be used for software analysis and “traditional” visual tests.
Our DBT QC Phantom consists of eight homogeneous slabs made from breast-equivalent material in 50/50 ratio of gland and adipose tissue (BR50/50). Optional swirled slabs are comprised of 100% adipose and 100% gland tissues in an approximate 50/50 ratio by weight. Three homogeneous slabs include imaging targets. Each slab measures 100 x 180 x 10 mm, allowing for the use of a combination of homogeneous and swirled slabs to test the influence of scatter radiation on image quality in more clinically-relevant, inhomogeneous conditions.
The phantom includes a positioning holder with magnetic fixation for repeatable alignment and positioning on DBT scanners.

Benefits:
Acceptance testing, daily and routine QC
Evaluate image quality and stability of DBT systems
Consistent and repeatable targets in homogeneous background
Optional complex background provides clinically relevant challenge for target detection
Developed to meet work-in-progress requirements of EUREF & AAPM TG245

Perform Image Quality Tests
Volume coverage missing tissues
Pixel Value Uniformity
Noise: SNR & SDNR
Resolution in X, Y and Z directions (MTF, ESF & PSF)
Geometric accuracy 3D
Artifact assessment
Visual detectability: specs, masses & fibers

Reference